CDRH is issuing this Early Alert to notify the public of a potentially high-risk device issue. The FDA will keep the public informed and update this web page as significant new information becomes available.
Affected Product
The FDA is aware that AVID Medical has issued a letter to affected customers recommending certain convenience kit be corrected prior to continued use by removing components affected by the Huons Co., Ltd. recall.
Affected devices:
| Model/Item Number (Kit) | Description (Tray Name) | Kit Lot Number | UDI |
|---|---|---|---|
| BMCN001-05 | CONTINUOUS NERVE BLOCK TRAY | 1703122 | 10809160455128 |
| BMCN001-05 | CONTINUOUS NERVE BLOCK TRAY | 1707656 | 10809160455128 |
| BMCN001-05 | CONTINUOUS NERVE BLOCK TRAY | 1708053 | 10809160455128 |
| DPGE002-02 | 18G SINGLE SHOT EPI TRAY | 1641713 | 10809160410325 |
| DPGE002-02 | 18G SINGLE SHOT EPI TRAY | 1641714 | 10809160410325 |
| DPGE002-02 | 18G SINGLE SHOT EPI TRAY | 1659697 | 10809160410325 |
| DPGE002-02 | 18G SINGLE SHOT EPI TRAY | 1668209 | 10809160410325 |
| PM415 | BASIC BIOPSY TRAY | 1652485 | 10809160322154 |
| PM415 | BASIC BIOPSY TRAY | 1656642 | 10809160322154 |
| PM415 | BASIC BIOPSY TRAY | 1667835 | 10809160322154 |
| PM415 | BASIC BIOPSY TRAY | 1676219 | 10809160322154 |
| PM415 | BASIC BIOPSY TRAY | 1682251 | 10809160322154 |
| PM415 | BASIC BIOPSY TRAY | 1691940 | 10809160322154 |
| VARV042-03 | UNIV IMAGING BIOPSY TRUCUSTOM | 1668365 | 10809160490822 |
| VARV042-03 | UNIV IMAGING BIOPSY TRUCUSTOM | 1668366 | 10809160490822 |
| VARV042-03 | UNIV IMAGING BIOPSY TRUCUSTOM | 1688166 | 10809160490822 |
| VARV042-03 | UNIV IMAGING BIOPSY TRUCUSTOM | 1692637 | 10809160490822 |
| VARV042-03 | UNIV IMAGING BIOPSY TRUCUSTOM | 1696930 | 10809160490822 |
What to Do
Do not use sodium chloride ampules included in affected kits. Check your stock for convenience kits containing sodium chloride ampules. Quarantine kits until new labels can be printed and affixed to all kits. All remaining components within the kits may continue to be used.
On August 10, AVID Medical sent all affected customers a letter recommending the following actions:
- Identify, Segregate, and Quarantine all impacted kits.
- Add warning labels to affected convenience kits to ensure all users are aware of the actions to take prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory.
- This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. If pre-printed labels are needed, please contact your sales representative.
- REMOVE AND DISCARD the Sodium Chloride, 0.9%, USP 10 mL Flush ampules prior to use.
Check this web page for updates. The FDA is currently reviewing information about this potentially high-risk device issue and will keep the public informed as significant new information becomes available.
Reason for Alert
AVID Medical stated that epidural kits may contain sodium chloride ampules manufactured by Huons Co., Ltd. and recalled by Spectra Medical Devices, LLC, due to concerns of product quality and sterility assurance in the Sodium Chloride Injection 0.9%, USP 10ml. The use of sodium chloride ampules may expose patients to viable microorganisms, endotoxins, and foreign matter that may trigger inflammatory responses and infections. This consideration is of particular importance in critically ill patients and when treatment is administered in close proximity to organs highly susceptible to injury — such as the spinal cord — where it may predispose to abscess formation and subsequent paralysis.
As of August 10, AVID Medical has reported no serious injuries and no deaths associated with this issue.
Device Use
Convenience kits contain various surgical instruments, dressings and/or other materials, and pharmaceutical components that are intended to be used during various surgical procedures.
Contact Information
Customers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact AVID Medical at complaints@owens-minor.com.
Additional FDA Resources
- Related Recalls
Unique Device Identifier (UDI)
The unique device identifier (UDI) helps identify individual medical devices sold in the United States from distribution to use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified more quickly, and as a result, problems potentially resolved more quickly.
- How do I recognize a UDI on a label?
- AccessGUDID database – Identify Your Medical Device
- Benefits of a UDI System
How do I report a problem?
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.